How do I change my KANE3500 let by, stabilisation & tightness intervals?
How do I change my KANE3500 let by, stabilisation & tightness intervals?
Your KANE3500 default interval times are:
1 minute Let by – TIME 01
1 minute Stabilisation – TIME 02
2 minute Tightness – TIME 03
You can change these intervals to any time up to 99 minutes as follows:
Press ON/OFF
to turn on your KANE3500
Press MENU
then Up / Down
arrow buttons to select PRESSURE
& press Enter / Arrow
button
Press Up / Down
to select REPORTS
& press Enter / Arrow
Press Up / Down
to select SETTINGS
& press Enter / Arrow
key
The display shows the Let by time:
TIME 01
TIME 00
Use Up / Down to select your time then Enter to move to the next digit & repeat - The cursor indicates which digit will change. Press Enter again to move to TIME 02
& 03
Repeat the above steps to change Stabilisation (TIME 02
) & Tightness (TIME 03
)
These times are stored in memory until you change them or remove the battery.