How do I change my KANE458 let by, stabilisation & tightness intervals?
Your KANE458 default interval times are:
1 minute Let by
1 minute Stabilisation
2 minutes Tightness
You can change these intervals up to 99 minutes as follows:
Rotate dial to
MENU
, use Up / Down buttons to chooseUNITS
, then pressENTER
Use Up / Down buttons to select
PRESSURE
, then pressENTER
Use Up / Down buttons to select
TIME
, then pressENTER
Use Up / Down buttons to select which interval to change, then press
ENTER
Use Up / Down buttons to change time then press
ENTER
Continue changing other intervals as above & / or rotate dial to exit
MENU
These times are stored in memory until you change them or remove the battery